| Earth resistance ranges | 200.0/2000Ω (Auto ranging) ±3%rdg±5dgt |
|---|---|
| Voltage ranges | AC: 5.0 - 300.0V (45 - 65Hz)±1%rdg±4dgt DC: ±5.0 - 300.0V±1%rdg±8dgt |
| Applicable standards | IEC 61010-1 CAT III 300V pollution degree 2 IEC 61010-031, 61557-1, -5 IEC 61326-1, 2-2, IEC 60529 (IP40) |
| Power source | Size AA alkaline battery × 2pcs |
| Dimensions | 232(L) × 51(W) × 42(D)mm |
| Weight | 220g approx. (including battery) |
| Included Accessories | 7248 (Test lead with Alligator clip and Flat test probe) 8072 (CAT II Standard prod) 8253 (CAT III Standard prod) 8017 (Extension prod long) 9161 (Carrying case) Instruction manual, LR6 (AA) × 2 |
| 4105A Soft Case Model | 4105A-H Hard Case Model |
|---|---|
![]() |
![]() |
| Measurement ranges | Earth Resistance: 0 - 20Ω/0 - 200Ω/0 - 2000Ω Earth Voltage [50,60Hz]: 0 - 200V AC |
|---|---|
| Accuracy | Earth Resistance: ±2%rdg±0.1Ω (20Ωrange) ±2%rdg±3dgt (200Ω/2000Ωrange) Earth Voltage: ±1%rdg±4dgt |
| Overload protection | Earth Resistance: 280V AC for 10 seconds across 2 of the 3 terminals Earth Voltage: 300V AC for 1 minute |
| Applicable standards | IEC 60529 IP54 |
| Safety Standard | IEC 61010-1 CAT III 300V Pollution Degree 2, IEC 61557 |
| Power source | R6 (AA) (1.5V) × 6 |
| Dimensions | 105(L) × 158(W) × 70(D)mm |
| Weight | 550g approx. |
| Included Accessories | 7095A (Earth resistance test leads) × 1set (red-20m, yellow-10m, green-5m) 8032 (Auxiliary earth spikes) × 1set 7127A (Simplified measurement probe) × 1set R6 (AA) × 6 9121 (Shoulder strap) Instruction Manual Carrying case : 9084(Carrying case[Soft]:4105A only) : 9165(Carrying case[Hard]:4105A-H only) |
| Optional Accessories | 7100A (Precision measurement cord set) |
| 4102A Soft Case Model | 4102A-H Hard Case Model |
|---|---|
![]() |
![]() |
| Measurement ranges | Earth Resistance: 0 - 12 Ω/0 - 120 Ω/0 - 1200 Ω Earth Voltage [50,60Hz]: 0 - 30V AC |
|---|---|
| Accuracy | Earth Resistance: ±3% of full scale Earth Voltage: ±3% of full scale |
| Overload protection | Earth Resistance: 276V AC for 10 seconds across 2 of the 3 terminals Earth Voltage: 276V AC for 1 minute |
| Applicable standards | IEC 61010-1 CAT III 300V Pollution Degree 2, IEC 61557 IEC 60529 IP54 |
| Power source | R6 (AA) (1.5V) × 6 |
| Dimensions | 105(L) × 158(W) × 70(D)mm |
| Weight | 600g approx. |
| Included Accessories | 7095A (Earth resistance test leads) × 1set (red-20m, yellow-10m, green-5m) 8032 (Auxiliary earth spikes) × 1set 7127A (Simplified measurement probe) × 1set R6 (AA) × 6 9121 (Shoulder strap) Instruction Manual Carrying case : 9084(Carrying case[Soft]:4102A only) : 9164(Carrying case[Hard]:4102A-H only) |
| Optional Accessories | 7100A (Precision measurement cord set) |