| AC A | 400A (0 - 400A) ±1%rdg±3dgt [50/60Hz] ±2%rdg±3dgt [40Hz - 1kHz] 2000A (0 - 1500A) ±1%rdg±3dgt [50/60Hz] ±3%rdg±3dgt [40Hz - 1kHz] 2000A (1500 - 2000A) ±3.0%rdg [50/60Hz] |
|---|---|
| AC V | 40/400/750V ±1%rdg±2dgt [50/60Hz] ±1.5%rdg±3dgt [40Hz - 1kHz] |
| DC V | 40/400/1000V ±1%rdg±2dgt |
| Ω | 400Ω/4k/40k/400kΩ ±1.5%rdg±2dgt |
| Continuity buzzer | buzzer sounds below 50±35Ω |
| Conductor size | Ø55mm max. |
| Frequency response | 40Hz - 1kHz |
| Output | Recorder : DC400mV against AC400A DC200mV against AC2000A |
| Safety Standard | IEC 61010-1 CAT III 600V , CAT II 1000V IEC 61010-2-031 IEC 61010-2-032 |
| Power source | R6 (AA) (1.5V) × 2 |
| Dimensions | 247(L) × 105(W) × 49(D)mm |
| Weight | 470g approx. |
| Included Accessories | 7107A (Test leads) 8201 (Output Plug) 9094 (Carrying Case) R6 (AA) × 2 Instruction Manual |
| Optional Accessories | 8008 (Multi-Tran) 7256 (Output cord) |


| AC A (RMS) | ||
|---|---|---|
| Range | 4.000/40.00/400.0A | |
| Accuracy | ±3%rdg±5dgt[45-500Hz] (At the center of the circle formed by the flexible sensor) |
|
| Crest factor (CF) |
Full scale CF<1.6, half scale<3.2 Effective input crest values are √2 times of the max values of each range. |
|
| Conductor size | Ø70mm max. | |
| Influence of Conductor position |
Additional ±2% (max.) depending on the distance from the center position | |
| Overload protection | 500A AC for 10 seconds | |
| Applicable standards | IEC 61010-1, IEC 61010-2-032 CAT Ⅳ 600V / CAT Ⅲ 1000V Pollution degree 2 IEC 61326-1(EMC) , IEC 60529 IP40 , EN50581(RoHS) |
|
| Operating temperature & humidity |
0 - +50℃,less than 80% RH (without condensation) | |
| Storage temperature & humidity |
-10 - +60℃,less than 70% RH (without condensation) | |
| Power source | R03 / LR03(AAA)(1.5V)×2 ※Continuous measuring time:approx.120hours (Auto power off:approx.15 minutes) |
|
| Dimensions | 120(L)×70(W)×26(D):Display unit φ5.5mm approx.:Cable diameter 1.8m approx.:Cable length |
|
| Weight | 200g approx. (including batteries) | |
| Included Accessories | 9174 (Carrying case) LR03(AAA)×2, Insruction manual |
|




| AC A | 40.00/400.0/1000A (Auto-ranging) ±1.5%rdg±5dgt (45 - 65Hz) ±2.0%rdg±5dgt (40 - 1kHz) |
|---|---|
| AC V | 4.000/40.00/400.0/600V (Auto-ranging) ±1.8%rdg±7dgt (45 - 65Hz) ±2.3%rdg±8dgt (65 - 500Hz) |
| DC V | 400.0mV/4.000/40.00/400.0/600V (Auto-ranging) ±1.0%rdg±3dgt (400mV range is excluded.) |
| Ω | 400.0Ω/4.000/40.00/400.0kΩ/4.000/40.00MΩ (Auto-ranging) ±2.0%rdg±4dgt (0 - 400kΩ) ±4.0%rdg±4dgt (4MΩ) ±8.0%rdg±4dgt (40MΩ) |
| Continuity buzzer | buzzer sounds below 50±30Ω |
| Conductor size | Ø33mm max. |
| Applicable standards | IEC 61010-1 CAT IV 300V / CAT III 600V Pollution degree 2 (AC A), CAT III 300V / CAT II 600V Pollution degree 2 (AC/DC V) IEC 61010-031, IEC 61010-2-032, IEC 61326 (EMC), EN 50581 (RoHS) |
| Power source | R03/LR03 (AAA) (1.5V) × 2 *continuous measuring time: approx. 120 hours (Auto power off: approx. 10 minutes) |
| Dimensions | 190(L) × 68(W) × 20(D)mm |
| Weight | Approx. 120g (including batteries) |
| Included Accessories | 7107A (Test leads), 9160 (Carrying case), R03 (AAA) × 2, Instruction Manual |
| Optional Accessories | 8008 (Multi-Tran) |
We have 43 guests and no members online